9611 sc Phosphate Low Range Analyzer, open chemistry, 4 channels, 24 VDC

9611 sc Phosphate Low Range Analyzer, open chemistry, 4 channels, 24 VDC
Product #: 9611.KTO.PL.D4U
Unit Price Contact Hach

90 days of continuous runtime

Only two liters of reagent are required for the analyzer to perform unattended for up to 90 days; twice as long as the Series 5000.

Save time on maintenance

The industry's only pressurized reagent delivery system eliminates the frequent maintenance associated with pumps.

Avoid downtime

Predictive diagnostic tools, including Hach’s proprietary Prognosys technology, warning LEDs, and high-visibility notification screens let you avoid unplanned downtime.

Verify easily with Hach Lab products so you don't waste time second-guessing

Grab Sample In and Out features allow quick analysis of a grab sample poured into the analyzer, and facilitate taking a sample out of the analyzer to verify in a lab test.

Specifications

Accuracy: LR model: ±4 µg/L or ±4% of reading, whichever is greater;
HR model: ±500 µg/L or ±5% of reading, whichever is greater
(Tested with Hach reagents)
Alarm: Four relays output; type: not powered SPDT relays, each rated at
5 A resistive, 240 VAC maximum
Connection: 18 to 14 AWG wire, 18 AWG stranded recommended
Application: Pure water / Power
Certifications: CE (EN 61326-1: 2006; EN 61010-1: 2010; EN 60529: 1991, +A1:2000 )
KC (EN 61326-1: 2006)
C-tick (EN 61326-1: 2006)
cETLus (UL 61010-1: 2012; NEMA 250: 2003; CSA C22.2 No 61010-1: 2012)
Controller Compatibility: SC200, SC1000
Dimensions (H x W x D): 804 mm x 452 mm x 360 mm
Enclosure rating: IP66 / NEMA 4X
Fitting: Sample line and sample bypass drain: 6 mm (¼-in.)

Air purge inlet: 6 mm (¼-in.) Chemical and case drains: 9.5 mm (3/8-in.)
Grab Sample: Grab Sample In and Grab Sample Out capability
Light source: Class 1M LED (light emitting diode) with a peak
wavelength of 810 nm
Lower Limit of Detection (LOD): LR model: 4 µg/L

HR model: 200 µg/L
Manual Languages: English
French
Spanish
B. Portuguese
Chinese
Japanese
Korean
Thai
German
Italian
C. Portuguese
Czech
Danish
Dutch
Polish
Swedish
Finnish
Bulgarian
Hungarian
Romanian
Lithuanian
Russian
Turkish
Slovak
Slovenian
Croatian
Greek
Estonian
Measuring Principle: Colorimetric
Measuring range: 4-5000 µg/L as PO4;
Mounting: Wall, panel or table
Number of Channels: 1, 2, 4; 4
Operating Humidity: 5 - 95 % non-condensing (indoor use only)
Operating temperature range: 5 - 45 °C
Outputs: 4 - 20 mA
Parameter: Low Range Phosphate
Pollution Degree: 2/II
Power requirements (Voltage): 24 VDC
Protection Class: I
Reagent Consumption: 2 L of each reagent every 90 days with 15 minute cycle time
Relays: Four; type: not powered SPDT relays, each rated at 5 A resistive, 240 VAC maximum
Repeatability: LR model: ±1%
HR model: ±500 µg/L or ±5% of reading, whichever is greater
Response time: Typically, 9.5 minutes at 25 °C (77 °F); changes with temperature
Sample Flow Rate: 55 - 300 mL/min
Sample Pressure: 2 - 87 psi (to Preset Pressure Regulator)
Sample Temperature: 5 - 50°C (41 - 122°F)
Storage conditions: -20 - 60 °C
Weight: 20 kg (45 lb) without reagents and standards, 36.3 kg (80 lb) with reagents