Chemistries                  603-2779 2080

CL17 Analyser

OBSOLETE ITEM

CL17 Analyser
Product #: 5440001-MY
Brand: HACH

OBSOLETE ITEM

Colorimetric chemistry - simple, fast, accurate

Fast automatic operation

Complete control - minimal maintenance

Rugged, lightweight enclosure

What's included?

CL17 Chlorine Analyser with 1 Month Supply of Free Chlorine Reagents, Maintenance kit, Installation Kit, and Manual. (The power cord is ordered separately.)

Specifications

Accessories: Installation Kit
Maintenance Kit
Accuracy: ± 5 % or ±0.03 mg/L as CL₂, whichever is greater
Air purge: 0.1 cfm (0.17 m³/h) instrument quality air at max. 20 psig (ca. 1.4 bar) with 1/4-inch OD tube, quick disconnect fitting
Alarm: Two alarms selectable for sample concentration alarm, analyzer system warning, or analyzer system shut-down alarm. Each is equipped with an SPDT relay with contacts rated for 5A resistive load at 230 V AC.
Certifications: Europe, CE Approved with:
EN 61326-1
CISPR 11
EN 61010-1
IEC 60529

North America:
UL 61010A-1
CAN/CSA C22.2 No. 1010.1-92
Communication Capabilities: 4-20 mA Output
Cycle Time: 2.5 min
Description: CL17 Chlorine Analyzer with Reagents for Free Chlorine
Dimensions (H x W x D): 454 mm x 314 mm x 179 mm
Display: LCD, 3½ inch digit measurement readout and six-character alphanumeric scrolling text line
Drain: ½ inch ID flexible hose, hose barb
Enclosure Construction: ABS plastic, two clear polycarbonate windows, IP62-rated with the gasketed door latched.
Enclosure rating: IP62
Includes: CL17 Chlorine Analyser with 1 Month Supply of Free Chlorine Reagents, Maintenance kit, Installation Kit, and Manual. (The power cord is ordered separately.)
Inlet: ¼" OD polyethylene tube, quick-disconnect fitting
Inlet Pressure: 1 - 5 psi (0.07 - 0.34 bar), 1.5 psi (0.103 bar) is optimum
Interferences: Other oxidizing agents such as bromide, chlorine dioxide, permanganate and ozone will cause a positive interference. Hexavalent chromium will cause a positive interference: 1 mg/L Cr6+ = approximately 0.02 mg/L as Cl₂. Hardness must not exceed 1,000 mg/L as CaCO3.
Light source: Class 1 LED (light emitting diode) with a peak wavelength of 520 nm; 50,000 hours estimated minimum life
Lower Limit of Detection (LOD): 0.03 mg/L (ppm)
Mounting: Wall mount
Operating Humidity: Up to 90 % at 40 °C (104 °F) maximum
Operating temperature range: 5 - 40 °C max.
Parameter: Chlorine, Free & Total
Power requirements (Amps): 2.5 A
Power requirements (Hz): 50/60 Hz
Power requirements (Voltage): 100 - 115/230 V AC
Pressure Range: Inlet Pressure to Sample Conditioning: 1.5 - 75 psi (0.1 - 5.2 bar)
Pressure Regulator Included?: No
Range: 0 - 5 mg/L Cl₂
Recorder Outputs: One 0/4-20 mA with an output span programmable over any portion of the 0 to 5 mg/L range. Recommended load impedance 3.6 to 500 ohms, 130 V isolation from earth ground.
Repeatability: 0.01 mg/L or 3.0 % , whichever is greater
Sample Flow Rate: 200 - 500 mL per minute minimum
Sample Temperature: 5 - 40 °C (41 - 104 °F)
Warranty: 12 months
Weight: 10.49 kg
What's included?: CL17 Chlorine Analyser with 1 Month Supply of Free Chlorine Reagents, Maintenance kit, Installation Kit, and Manual. (The power cord is ordered separately.)