Accuracy:
Free Chlorine: ± 3 % of the reference test* (DPD) at constant pH less than 7.2 (±0.2 pH unit)
Accuracy 2:
Free Chlorine: ± 10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
Accuracy 3:
Total Chlorine: ± 10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
Accuracy 4:
Total Chlorine: ± 20 % of the reference test** (DPD) at stable pH greater than 8.5
Cable Connection:
5 pin, M12 connector
Cable Length:
1 m (between gateway and sc controller)
Calibration Method:
1-point or 2-point (zero and slope) calibration
Certifications:
CE compliant for conducted and radiated emissions:
- CISPR 11 (Class A limits)
- EMC Immunity EN 61326-1 (Industrial limits)
Additional certifications when connected to an sc controller
Connection Drain Line:
1/2 in ID
Connection Sample Line:
1/4 in OD
Controller:
Panel only
Controller Compatibility:
SC200, SC1000, SC4500. All controllers sold separately
Description:
CLF10sc Free Chlorine Analyzer (Panel Only) with pHD Differential Sensor
Dimensions:
mm x 25 mm (7.68 in x 0.98 in)
Sensor Only: 195
Fitting:
USA
Includes:
CL10sc Panel, 1 m digital extension cable, Panel Manual, pHD Sensor Manual, Chlorine Sensor Manual.
Interferences:
Free chlorine: monochloramine, chlorine dioxide, ozone, and chalk deposits.
Total chlorine: chlorine dioxide, ozone and chalk deposits
Length:
Sensor: 195 mm
Lower Limit of Detection (LOD):
30 ppb (0.03 ppm)
Material:
Corrosion-resistant materials, fully-submersible
(stainless steel, PEEK, PVC, silicon rubber and polycarbonate)
Measurement method:
Reagentless, electrochemical, three-electrode amperometric system
Mounting:
Panel or Wall Mounting
Operating temperature range:
0 - 45 °C
Options:
pHD Differential Sensor
Parameter:
Free Chlorine
pH Monitoring Required?:
With differential pH sensor
pH Range:
4 - 9
Power requirements (Voltage):
None
Pressure Range:
0.5 bar, no pressure impulses and/or vibrations
Range:
0 - 20 ppm Cl₂
Repeatability:
30 ppb or 3%, whichever is greater
Response time:
Free Chlorine: 140 s or less for 90% change (T90)
at a stable temperature and pH
Response Time:
Total Chlorine: 100 s or less for 90% change (T90)
at a stable temperature and pH
Sample Flow Rate:
30 - 50 L/h (7.9 - 13.2 gal/hour), Optimal is 40 L/hour (10.5 gal/hour)
Sample Temperature:
5 - 45°C (41 - 113°F)
Sampling Time:
Continuous
Storage conditions:
-20 °C to 60 °C dry
Warranty:
12 months
Weight:
Approximately 5.5 kg (12 lbs)
Panel and empty panel-mounted components only
What's included?:
CL10sc Panel, 1 m digital extension cable, Panel User Manual
pHD electrode, PPS, for CL10sc (9181500), pHD Sensor User Manual
Digital extension cable, 1m (6122400)
Replacement Sensor for CLF10sc, SS Tip (8626200), Chlorine Sensor User Manual