Altitude:
3000 m maximum
Analogue Output Functional Mode:
Linear, PID
Communication (optional):
Analog:
Five 0-20 mA or 4-20 mA analog outputs on each analog output module
Up to two analog Input modules (0-20 mA or 4-20 mA). Each input module replaces a digital sensor input.
Digital:
Profibus DPV1 module
Modbus TCP
Profinet IO module
Ethernet IP module
Compatible Instruments:
Compatible Sensors and Analysers / Software Version (Release Year)
Amtax sc / V2.30 (2018) or higher
A-ISE sc / V1.02 or higher
AN-ISE sc / V1.08 (2013) or higher
N-ISE sc / V1.02 or higher
Nitratax clear sc, Nitratax eco sc, Nitratax plus sc / V3.13 (2013) or higher
NT3100sc/NT3200sc
Phosphax sc / V2.30 (2018) or higher
Phosphax sc LR/MR/HR / V1.01 (2018) or higher
TSS sc / V41.73 (2013) or higher
Solitax sc / V2.20 (2013) or higher
TU5300sc, TU5400sc / V1.34 (2017) or higher
SS7 sc (in Bypass) / V1.01 (2006) or higher
Ultraturb sc / V3.06 (2017) or higher
1720E / V2.10 (2006) or higher
Sonatax sc / V1.15 (2016) or higher
CL17sc / V2.7 (2019) or higher
CL10sc / V1.14 (2013) or higher
9184sc, 9185sc, 9187sc* / V2.03 (2013) or higher
Uvas plus sc / V3.01 (2017) or higher
LDO 2 sc* / V1.22 (2013) or higher
3798sc* / V2.03 (2013) or higher
3700sc + Inductive Conductive Digital 6120800 / V3.00 (2017) or higher
3422sc + Contacting Conductive Digital 6120700 / V3.00 or higher
3700 analog + Conductivity Module LXZ525.99.D0004
3400 analog + Conductivity Module LXZ525.99.D0004
pHD sc*, pHD-S sc / V3.10 (2016) or higher
1200-S sc* / V2.04 (2013) or higher
pHD analog + Digital Gateway 6120500 / V3.00 (2017) or higher
pHD analog + pH/ORP Module LXZ525.99.D0003
RC and PC analog sensor + Digital Gateway for conventional analog pH and ORP sensors 6120600 / V3.00 (2017) or higher
RC and PC analog + pH/ORP Module LXZ525.99.D0003
8362sc* / V3.00 (2017) or higher
Polymetron pH/ORP analog + Ultrapure pH/ORP Module LXZ525.99.D0007
Polymetron Conductivity analog + Ultrapure Conductivity Module LXZ525.99.D0006
GS1440 and GS2440EX Sensors H₂S
FP360 sc / V1 or higher
*Hardware Version1 of instrument is not supported
Compatible network technologies:
GSM 3G/4G (e.g. AT&T, T-Mobile, Rogers, Vodafone etc.)
CDMA (e.g. Verizon)
Compliance Certifications:
CE. ETL certified to UL and CSA safety standards (with all sensor types), FCC, ISED, KC, RCM, EAC, UKCA, SABS, C م (Morocco)
Conduit Openings:
½" NPT conduit
Description:
Microprocessor-controlled and menu-driven controller that operates the sensor
Dimensions:
½ DIN - 144 x 144 x 192 mm (5.7 x 5.7 x 7.6 in.)
Display:
3.5-inch TFT colour display with capacitive touchpad
Enclosure rating:
UL50E type 4X, IEC/EN 60529–IP 66, NEMA 250 type 4X
Metal enclosure with a corrosion-resistant finish
Indoor/Outdoor:
Outdoor installation in direct sunlight or UV radiation requires UV protection screen and/or sunroof
Installation category:
Category II
Material Enclosures:
Polycarbonate, aluminum (powder coated), stainless steel
Measurements:
Two device digital SC connectors
Mounting:
Wall, Pole, or Panel Mounting
Network connectivity:
LAN: Two Ethernet connectors (10/100 Mbps)
Cellular: External 4G
Wi-Fi
Operating temperature range:
-20 to 60 °C (-4 to 140 °F) (8 W (AC)/9 W (DC) sensor load)
-20 to 45 °C (-4 to 113 °F) (28 W (AC)/20 W (DC) sensor load)
Linear derating between 45 and 60 °C (-1.33 W/°C)
Outputs:
LAN + mA Output
Pollution Degree:
4
Power requirements (Voltage):
18-28 VDC; 2.5 A
Power supply:
Without plug 24 VDC C1D2
Protection Class:
I, connected to protective earth
Relays:
Two relays (SPDT);
Wire gauge: 0.75 to 1.5 mm² (18 to 16 AWG)
AC controller
Maximum switching voltage: 100 - 240 VAC
Maximum switching current: 5 A Resistive/1 A Pilot Duty
Maximum switching power: 1200 VA Resistive/360 VA Pilot Duty
DC controller
Maximum switching voltage: 30 VAC or 42 VDC
Maximum switching current: 4 A Resistive/1 A Pilot Duty
Maximum switching power: 125 W Resistive/28 W Pilot Duty
Sensor Input #1:
mA Input
Sensor Input #2:
mA Input
Software available:
Non-Claros + Embedded Prognosys
Storage conditions:
-20 to 70 °C, 0 - 95% relative humidity, non-condensing
USB Port:
Used for data download and software upload. The controller records approximately 20,000 data points for each connected sensor.
Warranty:
12 months
Weight:
1.7 kg (controller only, w/o modules)
What's included?:
SC4500 Controller, Non-Claros + Embedded Prognosys, without plug 24 VDC C1D2; includes mounting hardware