Accuracy:
Turbidity up to 1000 NTU:
without calibration < 5% of the measured value ±0.01 NTU
with calibration < 1% of the measured value ±0.01 NTU
Automatic Wiper:
Yes
Body Material:
Stainless Steel
Cable Length:
10 m (optional extension cables available)
Calibration Method:
Turbidity: Formazin or Stablcal Standard (at 800 NTU).
Suspended Solids: Sample specific, based on gravimetric TSS analysis with a correction factor procedure.
Certifications:
CE
Configurations:
Includes SC200 Controller and Insertion Mounting Kit
Controller:
SC200
Diameter Sensor:
60 mm
Flow:
Max. 3 m/s (the presence of air bubbles affects the measurement)
Includes:
Turbidity & Suspended Solids probe, controller, adapter, flange and insertion mounting, user manual
Installation style:
Ball Valve Insertion
Length:
315 mm
Maintenance Interval:
1 h/month
Material:
Optics Carrier and Sleeve: stainless steel 1.4571; Wiper Arm: stainless steel 1.4581; Wiper Rubber: silicone rubber (standard) Optional: FKM/FPM (LZX578); Wiper Shaft: stainless steel 1.4571
Max. cable length Sensor-to-Controller:
100 m
Measuring Principle:
Infrared Duo scattered light technique for color-independent
turbidity measurement Turbidity in accordance with
DIN EN 27027 / TS equivalent DIN 38414
Measuring range:
TSS-content: 0.001 mg/L - 50 g/L
Turbidity: 0.001 - 4000 NTU
Model:
Solitax inline sc
Mounting:
Insertion
Operating temperature range:
0 - 40 °C
Parameter:
Suspended Solids, Turbidity
Pressure Range:
Stainless steel insertion sensor: <6 bar or <60 m
Repeatability:
TSS content: < 3 %
Turbidity: < 1 %
Response Time T90:
1 - 300 s adjustable
Warranty:
12 months
Weight:
0.52 kg
What's included?:
Turbidity & Suspended Solids probe, controller, adapter, flange and insertion mounting, user manual