Accuracy:
5% of measured value + 0.2 mg/L (with standard solutions) NH4-N
Cable Length:
Standard: 10 m (33.8 ft)
Extension cables are available as an option in the following lengths:
5, 10, 15, 20, 30, 50 m (16.4, 33.8, 49.2, 65.6, 98.4, 164 ft)
Calibration Method:
With Cartrical plus technology: automatic import of factory calibration data from cartridge to probe by RFID*; 1 and 2 point matrix correction
Controller Compatibility:
SC200, SC1000, SC4500. All controllers sold separately
Data Transfer:
Current output, relays and bus communication via SC controller
Flow:
< 4 m/s max.
Installation Angle:
45 ° ± 15 ° (vertical in flow direction)
Length:
320 mm
Lower Limit of Detection (LOD):
0.2 mg/L NH₄–N
Material:
Cartridge: Stainless steel (1.4571), PVC, POM, ABS, NBR
Measurement method:
Potentiometric ion-selective electrodes for ammonium and potassium, reference system and temperature sensor
Measuring Interval:
Continuous
Measuring range:
0 - 1000 mg/L NH₄-N
Measuring uncertainty:
±5% of the measured value + 0.2 mg/L (with standard solution) NH4-N
Operating temperature range:
Air: -20 - 45 °C (-4 - 113 °F)
Parameter:
Ammonium
pH Range:
5 - 9 pH
Process Connection:
1 inch thread
Protection Class:
IP68
Repeatability:
±5 % of the measured value (with standard solution) NH4-N
Response time:
< 3 min
Sample Pressure:
0.3 bar max.
Sample Temperature:
2 - 40 °C (35.6 - 104 °F)
Sensor cartridge:
With Cartrical plus technology: compact housing containing calibrated electrodes for ammonium, potassium, reference system and temperature sensor, all calibrated to each other; typical lifetime 12 months
Sensor Immersion Depth:
0.3 - 3 m (1.0 - 10 ft)
Storage conditions:
5 - 40 °C sensor cartridge, -20 - 60 °C sensor
Warranty:
12 months
Weight:
2.38 kg
Wetted Materials:
Probe: Stainless Steel 1.4571; ASA + PC, silicon; PVC and PU;
Sensor cartridge: PVC; POM; ABS; Stainless Steel 1.4571; NBR
What's included?:
Ammonium probe, sensor cartridge, user manual