Would you like to use our Product Configurator tool to configure this product before adding it to your cart? If not, you may add it directly to your cart.
Legiolert detects L. pneumophila at 1 organism per 100 mL water sample.
- Detects Legionella pneumophila with equal or greater accuracy than traditional spread plate culture methods - Delivers definitive culture results up to 7 days faster than traditional culture methods without additional confirmation steps - Easy culture test setup and read significantly reduce laboratory work flow - Easy-to-use platform, similar to that of the Colilert Test - Received NF Validation by AFNOR Certification, reference number IDX 33/06-06/19
Potable Samples (10 mL protocol)
Step 1: Prepare 1:10 dilution of potable water sample in sterile diluent Step 2: Add reagent to water sample and shake to dissolve Step 3: Pour sample into Quanti-Tray/Legiolert Step 4: Seal using the Quanti-Tray Sealer PLUS and incubate for 7 days Step 5: Read results: Any brown and/or turbid wells are positive for Legionella pneumophila. Quanti-Tray/Legiolert provides Most Probable Number (MPN) results up to 2,272 per 100 mL sample
Non-potable Samples
Step 1: Add reagent to sterile diluent Step 2: Treat non-potable sample with Legiolert pretreatment Step 3: Add pre-treated sample to reagent mixture Step 4: Pour sample into Quanti-Tray/Legiolert Step 5: Seal using the Quanti-Tray Sealer PLUS and incubate for 7 days Step 6: Read results: Any brown and/or turbid wells are positive for Legionella pneumophila. Quanti-Tray/Legiolert provides Most Probable Number (MPN) results up to 2,272 per 100 mL sample